Product Overview
The Automatic Open Cup Flash Point Tester determines the open-cup flash point and fire point of petroleum products. It is designed for laboratories in the power, petroleum, chemical, commodity inspection, and scientific research sectors. The instrument complies with the test method requirements of GB/T 267-88 and GB/T 3536-2008.
The Automatic Open Cup Flash Point Tester determines the open-cup flash point and fire point of petroleum products. It is designed for laboratories in the power, petroleum, chemical, commodity inspection, and scientific research sectors. The instrument complies with the test method requirements of GB/T 267-88 and GB/T 3536-2008.
An ARM microprocessor, color LCD, and touchscreen interface provide convenient operation. Automated ignition, endpoint detection, result locking, printing, cooling, and power-off data retention help deliver accurate and repeatable measurements with minimal operator intervention.
How It Works
The sample is heated according to the temperature-rise profile specified in GB/T 267-88 and GB/T 3536-2008. As the expected flash point approaches, the instrument performs automatic ignition. Once a flash is detected, the measured value is automatically locked, displayed, and printed, and the heater is cooled automatically.
Key Advantages

Typical Applications
Suitable for routine quality control, product inspection, research, and laboratory testing in the power, petroleum, chemical, commodity inspection, and scientific research industries.

Model Selection Guide
| Parameter | Specification |
| Applicable standards | GB/T 267-88 and GB/T 3536-2008 |
| Measurement range | 40 to 400°C |
| Temperature sensor | Platinum resistance thermometer |
| Resolution | 0.1°C |
| Repeatability | ±2°C at flash points ≤150°C; ±4°C at flash points ≥150°C |
| Display | 7-inch color LCD touchscreen |
| Data storage | Up to 20,000 test results |
| Ignition method | Electric ignition |
| Cooling method | Forced-air cooling |
| Printer | Chinese character output, 40 lines |
| Self-test functions | Lifting mechanism, sweep arm, printer, and related components |
| Power consumption | ≤600 VA |
| Power supply | AC 220 V ±11 V, 50 Hz ±2.5 Hz |
| Operating temperature | 10 to 35°C |
| Operating humidity | ≤90% RH |
| Net weight | Approximately 12 kg |